Views: 2266

BWB Flash Flame Photometer

Product code : 1667895180

The BWB Flash is the first and only fully configurable up to four channel flame photometer that detects and displays all calibrated ions at the same time.The first simultaneous Flame Photometer with built in compressor and configurable element analysis to your requirements.

Quantity:
More information:
  • Nhà Sản Xuât
  • BMW Technologies

    Specifications:
    Sample rate
    2.5 - 4ml/min
    Tubing materials
    Silicone and Tygon®
    Required desk space for instrument
    101cm* (H) x 45cm (W) x 35cm (D)
    *For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
    Instrument size
    Flash Flame Photometer
    51cm (H) x 33cm (W) x 35cm (D) (20in x 13in x 14in)
    Shipping
    65cm (H) x 49cm (W) x 43cm (D) (25in x 19in x 17in)
    Weight
    BWB Flash – 10kg (22.1lbs)
    Shipping – 18kg (39.7 lbs)*
    *subject to vary based on accessories included
    Optimal range
    Single point calibration
    Na 0.1 – 60ppm
    K 0.05 – 100ppm
    Li 0.05 – 50ppm
    Ca 1.0 – 100ppm
     
    Multi Point Calibration (with dilution)
    Na 0.1 – 1000ppm
    K 0.05 – 1000ppm
    Li 0.05 – 1000ppm
    Ca 1.0 – 1000ppm
    Reproducibility
    <1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
    Specificity
    Na/K/Li = <1% to each other when equal in concentration at <100ppm
    Limit of detection (LOD) and limit of quantification (LOQ)
    LOD                             LOQ
    Na – 0.03ppm         Na – 0.1ppm
    K – 0.02ppm           K – 0.05ppm
    Li – 0.02ppm           Li – 0.05ppm
    Ca – 0.3ppm           Ca – 1ppm
    Time to stability
    Less than 15 seconds after sample is introduced into the flame
    Ions measured
    Simultaneous measurement of model specific elements
    Interfaces
    USB connectivity for supplied PC software
    .csv and .pdf generated reports and files via FP-PC software
    Recommended minimum warm up time based on ambient temperature 21ºC – 25 minutes
    Power requirements
    100V – 250V AC at 50 or 60Hz
    automatically selecting
    Fuel requirement*
    Propane, Butane or Natural Gas regulated to 19Bar. Flow rate of 0.4l/min
    *we recommend either Propane or Butane for optimum results
    Readout
    5” colour capacitive touch screen display